WEKO3
アイテム
Measurement Accuracy of Hand Dynamometers Used for Physical Fitness Testing
https://doi.org/10.14993/00000711
https://doi.org/10.14993/0000071102fac03b-f53a-4767-bd70-16ade72c8d54
| 名前 / ファイル | ライセンス | アクション |
|---|---|---|
|
|
|
| アイテムタイプ | 紀要論文 / Departmental Bulletin Paper(1) | |||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 公開日 | 2014-11-11 | |||||||||||||||||||||||||||||
| タイトル | ||||||||||||||||||||||||||||||
| タイトル | Measurement Accuracy of Hand Dynamometers Used for Physical Fitness Testing | |||||||||||||||||||||||||||||
| 言語 | en | |||||||||||||||||||||||||||||
| 言語 | ||||||||||||||||||||||||||||||
| 言語 | eng | |||||||||||||||||||||||||||||
| キーワード | ||||||||||||||||||||||||||||||
| 言語 | en | |||||||||||||||||||||||||||||
| 主題Scheme | Other | |||||||||||||||||||||||||||||
| 主題 | hand dynamometer | |||||||||||||||||||||||||||||
| キーワード | ||||||||||||||||||||||||||||||
| 言語 | en | |||||||||||||||||||||||||||||
| 主題Scheme | Other | |||||||||||||||||||||||||||||
| 主題 | measurement accuracy | |||||||||||||||||||||||||||||
| キーワード | ||||||||||||||||||||||||||||||
| 言語 | en | |||||||||||||||||||||||||||||
| 主題Scheme | Other | |||||||||||||||||||||||||||||
| 主題 | validation | |||||||||||||||||||||||||||||
| 資源タイプ | ||||||||||||||||||||||||||||||
| 資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||||||||||||||||||||||||||
| 資源タイプ | departmental bulletin paper | |||||||||||||||||||||||||||||
| ID登録 | ||||||||||||||||||||||||||||||
| ID登録 | 10.14993/00000711 | |||||||||||||||||||||||||||||
| ID登録タイプ | JaLC | |||||||||||||||||||||||||||||
| アクセス権 | ||||||||||||||||||||||||||||||
| アクセス権 | open access | |||||||||||||||||||||||||||||
| アクセス権URI | http://purl.org/coar/access_right/c_abf2 | |||||||||||||||||||||||||||||
| 著者 |
渡邊, 完児
× 渡邊, 完児
× 渡邊, 香織
× 吉岡, 昌美
|
|||||||||||||||||||||||||||||
| 抄録 | ||||||||||||||||||||||||||||||
| 内容記述タイプ | Abstract | |||||||||||||||||||||||||||||
| 内容記述 | The present study was conducted as a field survey to determine the actual degree of hand dynamometer validation performed in schools. The investigation consisted of a questionnaire survey of 43 schools (12 elementary, 6 junior high, and 25 senior high schools) that participated in a workshop for faculty members in the departments of health and physical education, together with actual validation tests for hand dynamometers in 7 of 43 schools (2 public elementary schools and 5 junior high schools), all 7 of which agreed to the testing. The dynamometer validation device of Takei Scientific Instruments Co., Ltd., was used for this study. The questionnaire responses indicated that 5 (12%) of 43 schools had performed hand dynamometer validation and 38 (88%) had not. Although this survey was small in scale, this finding suggests that many schools in Japan do not perform hand dynamometer validation, for various reasons. In the present study, in which an error of ±2.5 kg or more was considered large, the maximum deviation from the reference value was +8.0 kg in a hand dynamometer that showed a mean error of +6.5 kg in the test range of 10-70 kg. The present study has highlighted the issues associated with hand dynamometers used in physical fitness tests. It will be necessary in the future for teachers to raise consciousness about the validation testing of hand dynamometers in schools. | |||||||||||||||||||||||||||||
| 言語 | en | |||||||||||||||||||||||||||||
| 書誌情報 |
ja : 健康運動科学 en : Mukogawa Journal of Health and Exercise Science 巻 4, 号 1, p. 23-29, 発行日 2014-03-31 |
|||||||||||||||||||||||||||||
| 出版者 | ||||||||||||||||||||||||||||||
| 出版者 | 武庫川女子大学健康運動科学研究所 | |||||||||||||||||||||||||||||
| 言語 | ja | |||||||||||||||||||||||||||||
| ISSN | ||||||||||||||||||||||||||||||
| 収録物識別子タイプ | PISSN | |||||||||||||||||||||||||||||
| 収録物識別子 | 2185-338X | |||||||||||||||||||||||||||||
| 書誌レコードID | ||||||||||||||||||||||||||||||
| 収録物識別子タイプ | NCID | |||||||||||||||||||||||||||||
| 収録物識別子 | AA12576562 | |||||||||||||||||||||||||||||